Setup
3.4 Inputs and outputs
FST030
Function Manual, 04/2017, A5E35693328-AA
151
3.4.5.9
Digital input
3.4.5.9.1 Input function
Description
Sets the input functionality.
Setting Off; Reset totalizer 1; Reset totalizer 2; Reset totalizer 3; Reset all totalizers; Zero point
adjustment; Force outputs; Freeze process values
Default
Off
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Digital input → Input function
3.4.5.9.2 Forced value (Ch1)
Description
Forced value when forcing is active. Only applicable for HART option.
Setting
3.50 mA to 25.00 mA
Default
4.00 mA
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Digital input → Forced value
(Ch1)
3.4.5.9.3 Forced value (Ch2)
Description
Forced value when forcing is active.
Setting
3.50 mA to 25.00 mA
Default
4.00 mA
Current output
Setting
0.0 Hz to 12500.00 Hz
Default
1.00 Hz
Frequency output