Setup
3.2 Process values
FST030
Function Manual, 04/2017, A5E35693328-AA
61
3.2.6.1.4 Lower alarm limit
Description
Falling below this limit causes an alarm.
Setting
-9999997952 m/s to 9999997952 m/s
Default
-9999997952 m/s
Related
Path
Setup → Process values → LiquIdent → Alarm and warning limits → Lower alarm limit
3.2.6.1.5 Hysteresis
Description
Hysteresis that affects the alarm and warning limits.
Setting
0 m/s to +9999997952 m/s
Default
0.0 m/s
Related
Path Setup → Process values → LiquIdent → Alarm and warning limits → Hysteresis
3.2.6.2
Temperature slope factor
Description
Temperature slope factor - This factor is used in connection with the standard condition
temperature for calculating LiquIdent.
Setting
0.0 m/(s * K) to 8.0 m/(s * K)
Default
4.1 m/(s * K)
Related
Path
Setup → Process values → LiquIdent → Temperature slope factor
3.2.6.3
Pressure slope factor
Description
Pressure slope factor - This factor is used in connection with the standard condition pressure
to calculate LiquIdent.
Setting
-0.00003 m/(s * Pa) to 0.0 m/(s * Pa)
Default
-0.000004 m/(s * Pa)