Setup
3.2 Process values
FST030
74 Function Manual, 04/2017, A5E35693328-AA
3.2.12
Standard kinematic viscosity
3.2.12.1
Units
Description
Select unit for displayed standard kinematic viscosity. Unit is not applied at communication
interface.
Read only.
Setting
m
2
/s (square meters per second); cSt (1 centistoke = 1 mm
2
/s); St (stokes, 1 St = 1cm
2
/s)
Default
m
2
/s (square meters per second)
Related
Path
Setup → Process values → Standard kinematic viscosity → Units
3.2.12.2
Decimal places
Description
Sets the decimal places for the process value displayed in the local display.
Setting
Auto; No decimal place; 1 decimal place; 2 decimal places; 3 decimal places; 4 decimal
places; 5 decimal places; 6 decimal places
Default
2 decimal places
Related
Path
Setup → Process values → Standard kinematic viscosity → Decimal places
3.2.12.3
Alarm and Warning limits
3.2.12.3.1 Upper alarm limit
Description
Exceeding this limit causes an alarm.
Setting
-9999997952 m²/s to +9999997952 m²/s
Default
+9999997952 m²/s
Related
Path
Setup → Process values → Standard kinematic viscosity → Alarm and warning limits → Up-
per alarm limit