Setup
3.1 Sensor settings
FST030
38 Function Manual, 04/2017, A5E35693328-AA
3.1.7
User calibration
3.1.7.1
Slope
Description
A single point flow correction applied to volume flow equally at any flow rate.
Q = Q * Slope (e.g. Slope of 0.99 = -1% correction)
Setting
0.5 to 2.0
Default
1.0
Related
Path
Setup → Sensor settings → User calibration → Slope
3.1.7.2
Path 1 offset
Description
User calibrated zero offset for path 1.
Delta time offset correction value, which can be determined at zero flow.
Setting
-1250ns to 1250ns
Default 0.0
Related
Path
Setup → Sensor settings → User calibration → Path 1 offset
3.1.7.3
Path 2 offset
Description
User calibrated zero offset for path 2.
Delta time offset correction value, which can be determined at zero flow.
Setting
-1250ns to 1250ns
Default
0.0
Related
Path
Setup → Sensor settings → User calibration → Path 2 offset