3.4 Inputs and outputs
FST030
198 Function Manual, 04/2017, A5E35693328-AA
Related
Path
Setup → Inputs and outputs → Channel 4 - relay → On delay
3.4.8.18
Off delay
Description
Set the time delay between resetting alarm condition and de-activating the output.
Setting
0.00 s to 100.00 s
Default
0.00 s
Related
Path
Setup → Inputs and outputs → Channel 4 - relay → Off delay
3.4.9
Channel 5 - input
3.4.9.1
Process value
Description
Assignment of the current input value to a process value.
Setting
Off; Medium temperature (RTD); Sensor temperature (RTD); Auxiliary temperature (RTD)
Default
Off
Related
Path
Setup → Inputs and outputs → Channel 5 - Input → Process value
3.4.9.2
RTD configuration
3.4.9.2.1 RTD type
Description
Type of connected Resistance Temperature Detector (RTD).
Setting
100 ohms RTD; 500 ohms RTD; 1000 ohms RTD
Default
100 ohms RTD