Setup
3.4 Inputs and outputs
FST030
Function Manual, 04/2017, A5E35693328-AA
153
3.4.5.9.5 Delay time
Description
The debounce time used to delay the hardware input signal can be configured.
Setting 0.0 ms; 0.5 ms; 1 ms; 1.5 ms; 2 ms; 2.5 ms; 3 ms; 3.5 ms; 4 ms; 4.5 ms; 5 ms; 5.5 ms; 6
ms; 6.5 ms; 7 ms; 7.5 ms
Default
0.0 ms
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Digital input → Delay time
3.4.5.9.6 Polarity
Description
Logical polarity of discrete input signal.
Setting
Active high level; Active low level
Default
Active high level
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Digital input → Polarity
3.4.5.10
Current input
3.4.5.10.1 Process value
Description
Assignment of the current input value to a process value.
Read only.
Setting
Off; Pressure; Density; Kinematic viscosity; Medium temperature; Sensor temperature
(current input) Auxiliary temperature (current input)
Default
Off
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Current input → Process value
3.4.5.10.2 Change process value