Setup
3.1 Sensor settings
FST030
Function Manual, 04/2017, A5E35693328-AA
37
2.1.6.3.4 Zero point adjustment
Description
Zero point adjustment. Can only be changed via the Sensor Wizard.
Note
Important
Please stop all flow in the pipe and perform this function. The flow velocity is measured prior
and after the function has been applied.
Setting
5 to 300
Default
10
Related
Path
Setup → Sensor settings → Path settings → Path 2 → Zero point adjustment → Time duration
2.1.6.3.5 Serial number
Description
Serial number of sensor (clamp-on)
Setting
Default
Related
Path
Setup → Sensor settings → Path settings → Path 2 → Serial number
3.1.6.4
Necessary number of measuring paths
Description
List of measuring paths. If less than the specified number of paths are in a measuring state,
the meter will report an error.
3.1.6.5
Necessary number of measuring paths - settings
Setting
1 to 5
Default
5
Related
Path
Setup → Sensor settings → Path settings → Necessary number of measuring paths