Setup
3.1 Sensor settings
FST030
Function Manual, 04/2017, A5E35693328-AA
27
3.1.3.2
Disturbed flow profile compensation - settings
Setting
Enable; Disabled
Default
Disabled
Related
Path
Setup → Sensor settings → Upstream and downstream conditions → Disturbed flow profile
compensation
3.1.3.3
Type of upstream condition
Description
Compensation for upstream pipe geometry.
Setting
Straight run; Single elbow; Double elbow (in-plane); Double elbow (out-of-plane); Reducer;
Expansion
Default Straight run
Related
Path
Setup → Sensor settings → Upstream and downstream conditions → Type of upstream
condition
3.1.3.4
Upstream distance
Description
The distance to the upstream flow disturbance. The distance is given in pipe diameters.
Used for clamp-on flow profile compensation.
3.1.3.5
Upstream distance - settings
Setting
0 to 40
Default
0
Related
Path
Setup → Sensor settings → Upstream and downstream conditions → Upstream distance
3.1.3.6
Type of downstream condition
Description
Compensation for downstream pipe geometry.