Setup
3.1 Sensor settings
FST030
Function Manual, 04/2017, A5E35693328-AA
39
3.1.8
Multipoint calibration
3.1.8.1
Enable multipoint calibration
Description
Enable or disable the multipoint calibration.
Setting
Disabled; Enabled
Default
Disabled
Related
Path
Setup → Sensor settings → Multipoint calibration
3.1.8.2
Asymmetric calibration table
Description
Note
Access level: Expert user only
Only available for LUILOCAL_CURRENT_LUI_ACCESS_LEVEL == SUP AND
PID_ProtectC
alibrationTable == 0 AND PID_EnableUserCalibrationTable == 1
Usage of the calibration table:
Unidirectional: All 20 calibration points are used for positive flow calibration and are then
mirrored for negative flow.
Bidirectional: The first 10 calibration points are used for positive flow calibration and the
10 last calibration points are used for negative flow calibration.
Setting
Unidirectional; Bidirectional
Default
Unidirectional
Related
Path
Setup → Sensor settings → Asymmetric calibration table
3.1.8.3
Calibration point
Description
Calibration point that is to be transferred to the multipoint calibration table.
Setting
0.0 m
3
/s to 100000.0 m
3
/s
Default
0.0 m
3
/s