EasyManua.ls Logo

Siemens SITRANS F - Page 48

Siemens SITRANS F
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setup
3.2 Process values
FST030
46 Function Manual, 04/2017, A5E35693328-AA
3.2.1.6.2 Upper warning limit
Description
Exceeding this limit causes a warning.
Setting
-896.40 m
3
/h to 896.40 m
3
/h
Default
896.40 m
3
/h
Related
Path
Setup → Process values → Volume flow → Alarm and warning limits → Upper warning limit
3.2.1.6.3 Lower warning limit
Description
Falling below this limit causes a warning.
Setting
-896.40 m
3
/h to 896.40 m
3
/h
Default
896.40 m
3
/h
Related
Path Setup → Process values → Volume flow → Alarm and warning limits → Lower warning limit
3.2.1.6.4 Lower alarm limit
Description
Set Lower alarm limit. Alarm is triggered if measured value is below the limit.
Setting
-896.40 m
3
/h to 896.40 m
3
/h
Default
896.40m
3
/h
Related
Path
Setup → Process values → Volume flow → Alarm and warning limits → Lower alarm limit
3.2.1.6.5 Hysteresis
Description
Sets the hysteresis for alarm and warning limits. The hysteresis is the distance between the
limits for activation and deactivation of an alarm/warning.
Setting
0.00 m
3
/h to 896.40 m
3
/h
Default
0.0 m
3
/h

Table of Contents

Other manuals for Siemens SITRANS F

Related product manuals