Run Up / Coast Down Module
Displaying Acquired Data
Disp Order – (Forward or Reverse)
ForwardDisplay waterfall spectra from oldest (front of waterfall) to newest
(rear of waterfall).
ReverseDisplay spectra from newest (front of waterfall) to oldest (rear of
waterfall).
Figure 7 - 18.
Example Disp Order Settings of Forward (left) and Reverse (right).
Blanking (0% - 25%) - Removes a specified percentage from the noise floor of each
spectrum. Helps clarify the presence of machine orders or structural frequencies.
Figure 7 - 19.
Example Blanking Settings of 0%, 2%, and 5%.
After setting all Waterfall Disp. Settings, press the screen’s Display function
button to display the waterfall plot.
SKF Microlog - GX Series 7 - 19
User Manual