EasyManua.ls Logo

Keithley 4200-SCS - Page 120

Keithley 4200-SCS
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-14 Return to Section Topics 4200-900-01 Rev. K / February 2017
Section 3: Common Device Characterization Tests Model 4200-SCS User’s Manual
ITM timing
The ITM Timing window (see Figure 3-11) is used to set measurement speed and
the test mode. The ITM Timing window is opened by clicking the Timing button at
the top of the definition tab (see Figure 3-10).
Figure 3-11
ITM timing
Speed
There are four measurement speed settings:
Fast Good choice for quick measurements where noise and settling time are
not concerns.
Normal Provides a good combination of speed and low noise, and is the best
setting for most cases.
Quiet Best choice when you need the lowest noise and most accurate
measurements.
Custom Allows you to fine-tune the timing parameters to meet a particular need.
For details about this setting, refer to the Reference manual, Timing
window, page 6-132.
NOTE The above measurement speed selections can also be made from the speed drop-
down menu located at the top of the Definition tab (Figure 3-10).
A) Sweeping Mode selected
B) Sampling Mode selected

Table of Contents

Other manuals for Keithley 4200-SCS