EasyManuals Logo

Keithley 4200-SCS User Manual

Keithley 4200-SCS
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #264 background imageLoading...
Page #264 background image
3-158 Return to Section Topics 4200-900-01 Rev. K / February 2017
Section 3: Common Device Characterization Tests Model 4200-SCS User’s Manual
Figure 3-126
Slow single pulse—hardware connection
Figure 3-127
Example slow single pulse waveform graph
Model 4200-SCS
Instrument Slots
4205
PG2
4200
SCP2HR
V
D
V
G
INSTRUMENTS
SLOT
8
SLOT
7
SLOT
6
SLOT
5
SLOT
4
SLOT
3
SLOT
2
SLOT
1
4200
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4200
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4210
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
4210
SMU
SENSE LO
SENSE
FORCE
PA CNTRL
KEITHLEY
KEITHLEY
Channel 1
Channel 2
Ext
Trg
Ext
Clk
NOTE Use torque wrench to tighten SMA
connections to 8 inch-lbs.
Scope Card
Model 4200-PG2
Model 4200-SMU (1)
Model 4200-SMU (2)
AC+DC
Output
AC+DC Output
Substrate
Source
Gate
Drain
3-port
power
divider
1
2
S
SMA
Splitter
Tee
V
DD
V
g
CHANNEL 1
KEITHLEY
TRIGGER
OUT
CHANNEL 2
TRIGGER
IN
White SMA Cable (2m, 6ft)
(male-to-male)
1 of 6
= SMB Male to SMA Female Adapter
= BNC Male to SMA Female Adapter
White SMA Cable
(15cm, 6in)
(male-to-male)
800
600
400
200
0
020406080100
Time (
m
s)
t
r
t
f
V
g
= 1 to 2.2V
PW = 60ms
t
r
,t
f
= 5ms
I
d
degradation
PW = Pulse Width
Drain Current (
mA)

Table of Contents

Other manuals for Keithley 4200-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200-SCS and is the answer not in the manual?

Keithley 4200-SCS Specifications

General IconGeneral
BrandKeithley
Model4200-SCS
CategorySemiconductors
LanguageEnglish