EasyManua.ls Logo

Olympus 38DL PLUS - Figure 7-5 Measurement Screen for a Disbonded Oxide Layer

Olympus 38DL PLUS
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10004-01EN, Rev. D, November 2016
Chapter 7
98
To achieve accurate boiler tube and internal oxide scale thickness readings, remove
any oxide scale or coating from the external surface of the boiler tube.
The minimum internal oxide thickness that the 38DL PLUS can measure depends on
the material sound velocity but is approximately 0.254 mm (0.010 in.) with the M2017
transducer or 0.152 mm (0.006 in.) with the M2091 transducer. The 38DL PLUS only
displays the thickness of the steel boiler tube when the internal oxide scale thickness
is below the minimum measurement capability or when it is exfoliated (disbonded)
from the inside of the boiler tube.
Figure 7-5 on page 98 shows a waveform from a sample in which the internal oxide
layer is disbonded from the boiler tube. Because the oxide is disbonded, sound energy
does not travel into the internal oxide layer, and only a single reflected echo from the
inside of the boiler tube appears. When the internal oxide is thinner than 0.25 mm
(0.010 in.), the waveform looks almost identical. The steel/oxide interface is so close in
time to the echo from the inside of the oxide that it cannot be separated from the total
back-wall echo and will produce a single echo.
Figure 75 Measurement screen for a disbonded oxide layer
NOTE
Absence of a
steel/oxide echo
Tube thickness
Oxide layer thickness
Single echo from the
inside of boiler tube

Table of Contents

Other manuals for Olympus 38DL PLUS

Related product manuals