EasyManua.ls Logo

Olympus 38DL PLUS - Figure 12-4 Answering YES at the V-Path Calibration Prompt

Olympus 38DL PLUS
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10004-01EN, Rev. D, November 2016
Chapter 12
218
The thickness readings at this point may not be accurate because a proper zero offset
for the transducer has not yet been set.
c) Couple the transducer to the thinnest test block step.
d) If necessary, press [GAIN] to adjust the gain so that the instrument properly
detects the back-wall echo.
e) If necessary, adjust the extend blank so that the gage is making proper
detection on the back-wall signal (see “Blanking Adjustments in Manual
Echo-to-Echo Detection Mode” on page 83).
f) Ensure that the instrument properly detects echoes on all test block steps.
4. Activate the V-path function (see “Activating the V-Path Function” on page 216
for details).
5. Couple the transducer to a thick sample representing the thickest material that
you will measure.
6. Press [CAL VEL].
7. In the help text bar, select YES to the VPath Calibration? prompt (see Figure 12-4
on page 218).
Figure 124 Answering YES at the Vpath calibration prompt
8. While the transducer is coupled to the thick sample and getting a steady thickness
reading, press [ENTER].
9. In the VPATH CALIBRATION screen, edit the point 1 value to match the known
thickness (see Figure 12-5 on page 219).
NOTE

Table of Contents

Other manuals for Olympus 38DL PLUS

Related product manuals