DMTA-10004-01EN, Rev. D, November 2016
Dual Element Transducer Setups
219
Figure 125 Editing point 1 of the Vpath calibration
10. Press [CAL VEL].
11. Couple the transducer to the next thickest sample to get a steady thickness
reading, and then press [ENTER].
12. In the VPATH CALIBRATION screen, edit the value for the next point to match
the known thickness.
13. Repeat steps 10 to 12 for the other thicknesses. You need a minimum of three
points and a maximum of ten V-path calibration points.
14. When all known thickness values from all calibration points have been entered,
press [MEAS].
15. At this point you can either press [CAL VEL] to view the V-path Calibration table.
OR
Save the V-path calibration as a custom dual element transducer setup:
a) Press [SAVE/SEND].
b) In the SAVE SETUP screen, enter the desired custom setup file name (see
Figure 12-6 on page 220).