DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
153
To view an inserted or an appended cell in the DB grid
1. Activate and configure the DB grid (see “Activating and Configuring the DB
Grid” on page 147).
2. Press [ID#], and then use the arrow keys to move to the desired shaded grid cell.
3. Press [ZOOM].
The grid changes to a linear view and the inserted or appended ID number is
displayed (see Figure 10-15 on page 153).
Figure 10‑15 Example of a zoomed inserted cell
4. Press [ZOOM] again to return to the normal DB grid view.
10.5 Configuring AVG/MIN Measurements
You can use the AVG/MIN measurement function to store the average or the
minimum measurement of several successive thickness readings to the datalogger.
To configure the AVG/MIN measurement option
1. From the measurement screen, press [SETUP MENU].
2. In the menu, select AV G / M I N.
3. In AV G / M I N screen (see Figure 10-16 on page 154):
a) Set AVG/MIN ENABLE to ON to activate the average/minimum
measurement function.
b) Set MODE to AV E R A G E to calculate the average of the sequential thickness
readings or MINIMUM to calculate the minimum of the sequential thickness
readings.
c) Set # OF READINGS to set the number of thickness values in the sequence (2,
3, or 4).