Chapter 11
206
Figure 1134 The FILE COMPARISON report result screen
d) Review the maximum wall loss location list and the maximum wall growth
location list.
e) Select CANCEL to return to the measurement screen or NEW REPORT to
generate another report.
7. In the ALARM SUMMARY screen:
a) Select the file for which you want to generate the report.
b) Select REPORT.
The report page of the ALARM SUMMARY screen opens with the first low
alarm location ID highlighted (see Figure 11-35 on page 207).