Figure 126 Saving the setup
c) In the SAVE TO list, select a storage location.
DEUSERX (where X = 1 to 10) represents an empty location. Selecting a custom
setup location that already has a stored custom setup deletes and replaces the
contents with the new custom dual element transducer setup.
d) Select SAVE to save the custom setup.
e) In the ACTIVE screen, review the parameters.
f) Press [MEAS] to return to the measurement screen.
The new custom transducer is now the active setup and can be recalled from
the list of custom dual element transducer setups.