EasyManuals Logo

Olympus 38DL PLUS User Manual

Olympus 38DL PLUS
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #229 background imageLoading...
Page #229 background image
DMTA-10004-01EN, Rev. D, November 2016
Dual Element Transducer Setups
217
When using generic dual element transducers, you need to determine the exact
thickness range over which you will measure. The minimum and maximum
thicknesses are highly dependent on:
Frequency of the transducer
Material being tested
Roof angle of the transducer
Delay material in the transducer
Signal noise in the transducer
Olympus cannot guarantee the accuracy or performance with non-D79X transducers.
The user assumes all responsibility for assuring that the generic dual element
transducer is performing properly and is suitable for the application.
To build a V-path correction for a nonstandard dual element transducer
1. Ensure that the transducer is not coupled to the test material and that there is no
couplant on the tip to the transducer.
2. Press [2nd F], [CAL ZERO] (Do ZERO) to perform the transducer zero
calibration.
3. To ensure that the instrument properly detects echoes:
a) Press [RANGE] until the range value is higher than the maximum thickness
of the test block.
b) Couple the transducer to the thickest test block step, and then press [GAIN]
to adjust the gain so that the instrument properly detects the back-wall echo,
and then ensure that there are no noise signals above 20 % of the screen
height.
IMPORTANT
CAUTION

Table of Contents

Other manuals for Olympus 38DL PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus 38DL PLUS and is the answer not in the manual?

Olympus 38DL PLUS Specifications

General IconGeneral
Operating Temperature-10 °C to 50 °C (14 °F to 122 °F)
Data LoggerYes
InterfaceUSB, RS-232
Thickness Range0.08 mm to 635 mm (0.003 in. to 25 in.) depending on material and transducer
DisplayBacklit LCD
Battery TypeThree alkaline or NiMH AA cells
PulserAdjustable square wave pulser
Receiver Bandwidth0.5 MHz to 20 MHz
LanguagesEnglish, French, German, Spanish, Italian, Russian, Japanese, Chinese
Resolution0.001 mm or 0.0001 in.
Transducer Frequency2 MHz to 30 MHz
MaterialPlastic housing

Related product manuals