DMTA-10004-01EN, Rev. D, November 2016
Software Options
103
To use the multi-measurement with the soft contact mode
1. Create and store custom setups for the sagittal height and lens thickness
measurements (see “Creating a Setup for Nonstandard Dual Element
Transducers” on page 212).
2. Ensure that the multi-measurement software option is activated (see “Activating
Software Options” on page 92).
3. Press [SETUP MENU].
4. In the menu, select MULTI.
5. In the MULTI screen (see Figure 7-9 on page 104):
a) Set MULTI ENABLE to ON.
b) Set MULTI MODE to SOFT CONTACT.
c) Set DISPLAY MODE to WAVEFORM or LARGE FONT (see Figure 7-8 on
page 102).
d) Set SAVE/SEND KEY to one of the following options to determine the
operation when you press [SAVE/SEND]:
ACTIVE: Saves only the active measurement (the highlighted thickness)
to the datalogger.
AUTOINCR ACTIVE: Saves the active measurement to the datalogger,
and then changes the active setup to the next setup in the
multi-measurement list. This option allows you to easily save all
multilayer measurement by repeatedly pressing [SAVE/SEND].
e) In SGTTL HT, select the custom setup for the sagittal height measurement.
f) In LENS THK, select the custom setup for the lens thickness measurement.
g) In PDSTL DIA, enter the diameter of the pedestal being used.
h) Press [MEAS] to return to the measurement screen with the
multi-measurement function active in the soft contact mode.