DMTA-10004-01EN, Rev. D, November 2016
Chapter 10
146
3. Press [SAVE/SEND] while the minimum or maximum thickness reading is
displayed.
To save the minimum or maximum A-scan along with the frozen B-scan
1. Set BSCAN FREEZE MODE to SHOW MINIMUM or SHOW MAXIMUM.
The gage displays the minimum or maximum thickness reading with the
corresponding waveform.
2. While the B-scan is running, press [FREEZE] to enter the review mode.
3. Press [2nd F], [SAVE/SEND] (WAVEFORM) while the minimum or maximum
thickness reading is displayed.
To save the entire B-scan (B-scan history)
1. While the B-scan is running or when it is frozen, press [2nd F], [SAVE/SEND]
(WAVEFORM).
The Save BScan history? message appears on the help text bar.
2. Select Yes or No:
Yes
Saves the entire B-scan history including the Minimum or Maximum
thickness with corresponding waveforms.
No
Saves the current B-scan screen and the waveform.
When a B-scan screen is saved to the datalogger, the gage saves the thickness values
for the 600 data points appearing on the display. All thickness values on a saved
B-scan can be reviewed during ID Review. Recall the saved B-scan and use the []
and [] keys to review each thickness reading.
You can save a maximum of 10000 thickness readings to a B-scan. When the
maximum number of thickness points reaches 10000, the gage prompts you to save
the B-scan history or to reset the B-scan without saving.
NOTE
NOTE