EasyManuals Logo

Orolia SecureSync 2400 User Manual

Default Icon
589 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #318 background imageLoading...
Page #318 background image
Under the System tab, you can set Notifications for Minor and Major Alarms/Clearances.
The temperature readouts used for the Alarms are generated by the CPU temperature
sensor.
Also, you can set the temperature threshold value for Minor/Major alarms, and define a
retry value by determining how many readings (1/min.) the temperature must exceed the
threshold value before an alarm/clearance is triggered.
The default temperature threshold value for both Minor, and Major Alarms is 100°C. With
simultaneous alarm triggerings, the Major Alarm will override the Minor Alarm, i.e. you will
be notified only about the Major Alarm. If you want to be notified early about a rise in tem-
perature, a recommended setting for the Minor Alarm temperature would be 90°C. Please
302
CHAPTER 4 SecureSync 2400 User Manual Rev. 2
4.5 Quality Management

Table of Contents

Other manuals for Orolia SecureSync 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Orolia SecureSync 2400 and is the answer not in the manual?

Orolia SecureSync 2400 Specifications

General IconGeneral
BrandOrolia
ModelSecureSync 2400
CategoryRecording Equipment
LanguageEnglish