EasyManua.ls Logo

R&S SMA100B - How to Generate a Pulse Modulated Signal; How to Generate an Amplitude Modulated Signal

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Analog modulations
R&S
®
SMA100B
124User Manual 1178.3834.02 ─ 09
AM
Label for the amplitude modulation settings in the overview.
Remote command:
n.a.
FM/PhiM
Selects the modulation signal to be assigned to the output. You can assign the fre-
quency or phase modulated signal.
Remote command:
n.a.
5.5 How to generate an amplitude modulated signal
The following examples use the internal LF generator.
To set the frequency and level of the RF signal
1. Press [Preset] to start from a defined state.
2. In the status bar, set "Frequency = 2 GHz".
3. Set "Level = -20 dBm".
To configure the modulation signal (LF generator)
1. Select "Modulation" > "Modulation Source".
2. In the "Sources" tab, select "Shape > Sine".
3. Set "Frequency = 20 kHz".
The period is calculated automatically and indicates "Period = 50.00 µs".
To configure the amplitude modulation settings
1. In "Modulation" dialog, select the "AM" tab.
2. Set "Source = LF Generator1"
3. Set "Depth = 30%"
4. Select "State = I" to activate the modulation.
5. Set "Level > RF ON" to enable signal output.
5.6 How to generate a pulse modulated signal
The following example uses the internal pulse generator.
How to generate a pulse modulated signal