Improving level performance
R&S
®
SMA100B
230User Manual 1178.3834.02 ─ 09
Performs a zeroing procedure before acquiring the user correction data to improve
precision.
No signal is applied to the sensor during zeroing. RF output is temporarily switched
off during that time.
When unchecked, the zeroing procedure is skipped. However, the RF signal level
might be blanked shortly. This setting is useful if blanking of RF is undesirable or
the absence of power at the sensor cannot be guaranteed.
●
"Execute"
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
Remote command:
[:SOURce<hw>]:CORRection:ZERoing:STATe on page 639
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
on page 638
Used SMAB Settings For Measurement
Displays the settings relevant for the measurement.
"Modulation"
Indicates the modulation state
"Amplitude"
Displays the currently set level.
Remote command:
[:SOURce<hw>]:POWer:POWer on page 687
Remote command:
n.a.
8.3.4 Import/export list files
Access:
1. Select one of the following:
● "Sweep" > "List mode".
● "Level" > "User Correction".
● "Modulation > Pulse Modulation > Pulse Generator > Pulse Mode = Train".
User correction