Network operation and remote control
R&S
®
SMA100B
396User Manual 1178.3834.02 ─ 09
larly since faulty commands from the controller to the instrument are recorded there as
well.
13.4.8 Reset values of the status reporting system
The following table contains the different commands and events causing the status
reporting system to be reset. None of the commands, except of *RST and
SYSTem:PRESet affect the functional instrument settings. In particular, DCL does not
change the instrument settings.
Table 13-7: Resetting the status reporting system
Event Switching on supply
voltage
Power-On-Status-
Clear
DCL,
SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYSTem:
PRESet
STATus:
PRESet
*CLS
Effect 0 1
Clear STB, ESR - Yes - - - Yes
Clear SRE, ESE - Yes - - - -
Clear PPE - Yes - - - -
Clear error queue Yes Yes - - - Yes
Clear output buffer Yes Yes Yes 1) 1) 1)
Clear command processing
and input buffer
Yes Yes Yes - - -
1) The first command in a command line that immediately follows a <PROGRAM MESSAGE TERMINA-
TOR> clears the output buffer.
13.5 Remote access settings
This section outlines the settings required for accessing and configuring the provided
remote control interfaces. It includes network settings, access addresses, emulation
settings for using the command sets of other generators, and the access whith smart
devices.
About instrument emulations
You can remotely control the R&S SMA100B using the command set of another signal
generator. With this function you can, for example, replace a signal generator with an
R&S SMA100B in an automated test setup, without adjusting the command scripts
used.
The R&S SMA100B also covers command sets of Rohde & Schwarz signal genera-
tors. To achieve optimal compatibility when replacing an instrument, we recommend
that you select the emulation command set for the corresponding signal generator.
Remote access settings