Remote control commands
R&S
®
SMA100B
654User Manual 1178.3834.02 ─ 09
// activate the LF output and select the LF1 as signal source
// configure the LF output signal
SOURce1:LFOutput1:FREQuency:MODE CW
SOURce1:LFOutput1:STATe 1
SOURce1:LFOutput1:SOURce LF1
SOURce1:LFOutput1:INTernal:VOLTage 1
SOURce1:LFOutput1:OFFSet 0.001
LF generator settings............................................................................................ 654
LF sweep settings................................................................................................. 661
14.16.6.1 LF generator settings
With the commands described in this section, you can configure the LF signal source.
[:SOURce]:LFOutput<ch>:BANDwidth?............................................................................654
[:SOURce]:LFOutput<ch>:FREQuency.............................................................................655
[:SOURce<hw>]:LFOutput<ch>:PERiod?..........................................................................655
[:SOURce<hw>]:LFOutput:FREQuency:MANual............................................................... 655
[:SOURce<hw>]:LFOutput:FREQuency:MODE................................................................. 656
[:SOURce<hw>]:LFOutput:FREQuency:STOP.................................................................. 656
[:SOURce<hw>]:LFOutput:FREQuency:STARt..................................................................656
[:SOURce]:LFOutput<ch>[:STATe]................................................................................... 657
[:SOURce]:LFOutput<ch>:INTernal:VOLTage.................................................................... 657
[:SOURce]:LFOutput:OFFSet.......................................................................................... 657
[:SOURce]:LFOutput<ch>:SOURce................................................................................. 658
[:SOURce]:LFOutput<ch>:SOURce:PATH........................................................................ 658
[:SOURce]:LFOutput:VOLTage........................................................................................ 658
[:SOURce<hw>]:LFOutput<ch>:SHAPe............................................................................658
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:DCYCle....................................................659
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:PERiod.....................................................659
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:WIDTh......................................................659
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:FALL.................................................... 660
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:HIGH....................................................660
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod................................................. 660
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:RISE.................................................... 660
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod..................................................661
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE..................................................... 661
[:SOURce]:LFOutput<ch>:BANDwidth?
Queries the bandwidth of the external LF signal.
Return values:
<Bandwidth> BW0M2 | BW10m
*RST: BW10m
Example:
LFO:BAND?
// BW10m
// the bandwidth of the externally supplied LF signal is 10 MHz