Remote control commands
R&S
®
SMA100B
658User Manual 1178.3834.02 ─ 09
[:SOURce]:LFOutput<ch>:SOURce <Source>
Determines the LF signal to be synchronized, when monitoring is enabled.
Parameters:
<Source> LF1 | LF2 | NOISe | AM | FMPM | EXT1 | | EXT2
LF1|LF2
Selects an internally generated LF signal.
NOISe
Selects an internally generated noise signal.
EXT1|EXT2
Selects an externally supplied LF signal
AM
Selects the AM signal.
FMPM
Selects the signal also used by the frequency or phase modula-
tions.
*RST:
LF1
Example: See Example "Configuring the LF generator" on page 653.
Manual operation: See "Source" on page 121
[:SOURce]:LFOutput<ch>:SOURce:PATH <SourPath>
Determines the path of the LF output source.
Parameters:
<SourPath> A | B
*RST: A
Example:
LFO:SOUR:PATH?
Queries the currently set path for the LF output signal source.
[:SOURce]:LFOutput:VOLTage <Voltage>
Sets the voltage of the LF output.
Parameters:
<Voltage> float
Range: dynamic (see data sheet)
Increment: 0.001
*RST: 1
Example:
SOURce:LFOutput:VOLTage 1.5
[:SOURce<hw>]:LFOutput<ch>:SHAPe <Shape>
Selects the waveform shape of the LF signal.
SOURce subsystem