EasyManua.ls Logo

R&S SMA100B - Execute Single Sweep; Reset Sweep; Show Connector; Trigger Slope

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
List and sweep mode
R&S
®
SMA100B
193User Manual 1178.3834.02 ─ 09
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl on page 703
RF level sweep:
[:SOURce<hw>]:SWEep:POWer:DWELl on page 701
RF combined sweep:
[:SOURce<hw>]:SWEep:COMBined:DWELl on page 708
LF frequency sweep:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl on page 661
Trigger Slope
For "Mode = Extern Step/Single", selects the polarity of the active slope of an applied
instrument trigger.
Trigger signal is expected at the Inst Trig connector.
"Positive"
The rising edge of the trigger signal triggers the instrument.
"Negative"
The falling edge of the trigger signal triggers the instrument.
Remote command:
[:SOURce]:INPut:TRIGger:SLOPe on page 652
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Reset Sweep
Resets a sweep.
With the next trigger event, the sweep starts at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 710
Execute Single Sweep
In "Mode = Single", starts a sweep manually.
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute on page 705
:TRIGger<hw>:FSWeep[:IMMediate] on page 744
RF level sweep:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 705
:TRIGger<hw>:PSWeep[:IMMediate] on page 744
RF combined sweep:
[:SOURce<hw>]:SWEep:COMBined:EXECute on page 708
LF frequency sweep:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute on page 662
:TRIGger<hw>:LFFSweep:IMMediate on page 744
General:
:TRIGger<hw>[:SWEep][:IMMediate] on page 744