Remote control commands
R&S
®
SMA100B
744User Manual 1178.3834.02 ─ 09
:TRIGger<hw>:FPSWeep:SOURce <FpTrigSource>
Selects the trigger source for the combined RF frequency / level sweep.
The parameter names correspond to the manual control. If needed, see table
Table 14-1 for selecting the trigger source with SCPI compliant parameter names.
Parameters:
<FpTrigSource> AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO|IMMediate
Executes the combined RF sweep automatically. In this free-run-
ning mode, the trigger condition is met continuously. I.e. as soon
as a sweep is completed, the next one starts immediately.
SINGle|BUS
Executes one complete sweep cycle triggered by the GPIB com-
mands [:SOURce<hw>]:SWEep:COMBined:EXECute or
*TRG. The mode has to be set to [:SOURce<hw>]:SWEep:
COMBined:MODE > AUTO.
EXTernal
An external signal triggers the sweep.
EAUTo
An external signal triggers the sweep. As soon as one sweep is
finished, the next sweep starts. A second trigger event stops the
sweep at the current frequency and level value pairs, a third trig-
ger event starts the trigger at the start values, and so on.
*RST:
No enum range assigned!
Manual operation: See "Mode" on page 189
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
●
FSWeep - RF frequency
●
PSWeep - RF level
●
LFFSweep - LF frequency
●
SWEep - all sweeps
Effective in the following configuration:
●
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING
●
SOUR:SWE:FREQ|POW:MODE AUTO or SOUR:LFO:SWE:[FREQ:]MODE AUTO
Alternatively, you can use the IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
TRIG
Starts all active sweeps.
TRIGger subsystem