EasyManua.ls Logo

R&S SMA100B - Page 743

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
743User Manual 1178.3834.02 ─ 09
LFFSweep - LF frequency
PSWeep - RF level
SWEep - all sweeps
The source names of the parameters correspond to the values provided in manual
control of the instrument. They differ from the SCPI-compliant names, but the instru-
ment accepts both variants.
Use the SCPI name, if compatibility is an important issue. Find the corresponding
SCPI-compliant commands in Cross-reference between the manual and remote con-
trol.
Setting parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO [IMMediate]
Executes a sweep automatically.
In this free-running mode, the trigger condition is met continu-
ously. I.e. when a sweep is completed, the next one starts imme-
diately.
SINGle [BUS]
Executes one complete sweep cycle.
The following commands initiate a trigger event:
*TRG on page 454
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
:TRIGger<hw>[:SWEep][:IMMediate], :TRIGger<hw>:
PSWeep[:IMMediate] and :TRIGger<hw>:FSWeep[:
IMMediate].
Set the sweep mode with the commands:
[:SOURce<hw>]:SWEep:POWer:MODEAUTO|STEP
[:SOURce<hw>]:SWEep[:FREQuency]:MODEAUTO|STEP
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE
AUTO|STEP
In step mode (STEP), the instrument executes only one step.
EXTernal
An external signal triggers the sweep.
EAUTo
An external signal triggers the sweep. When one sweep is fin-
ished, the next sweep starts.
A second trigger event stops the sweep at the current frequency,
a third trigger event starts the trigger at the start frequency, and
so on.
*RST:
AUTO
Example: See Example "Setup an LF sweep" on page 652
Usage: Setting only
TRIGger subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals