Improving level performance
R&S
®
SMA100B
267User Manual 1178.3834.02 ─ 09
Mesial
Sets the medial reference level in terms of percentage of the overall pulse level (power
or voltage related). This level is used to define the pulse width (τ) and pulse period.
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
REFerence on page 509
:TRACe<ch>[:POWer]:SWEep:PULSe:THReshold:POWer:REFerence
on page 543
Proximal
Sets the lower reference level in terms of percentage of the overall pulse level (power
or voltage related).
The proximal power defines the start of the rising edge and the end of the falling edge
of the pulse.
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
LREFerence on page 509
:TRACe<ch>[:POWer]:SWEep:PULSe:THReshold:POWer:LREFerence
on page 543
Pulse data notifications settings
Access:
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Time".
4. Select "Configure Time Mode"
Using power sensors