Parameters:
<HReference> float
Range: 0.0 to 100.0
Increment: 0.01
*RST: 90.0
Example:
TRAC2:SWE:PULS:THR:POW:HREF?
Queries the upper reference level of trace 2.
Manual operation: See "Distal" on page 266
:TRACe<ch>[:POWer]:SWEep:PULSe:THReshold:POWer:LREFerence
<LReference>
Queries the lower medial threshold level of the overall pulse level. The proximal power
defines the start of the rising edge and the end of the falling edge of the pulse.
Note: This parameter is only avalaible in time measurement mode and R&S NRP-Z81
power sensors.
Parameters:
<LReference> float
Range: 0.0 to 100.0
Increment: 0.01
*RST: 10.0
Example:
TRAC:SWE:PULS:THR:POW:LREF?
Queries the medial threshold level of trace 1.
Manual operation: See "Proximal" on page 267
:TRACe<ch>[:POWer]:SWEep:PULSe:THReshold:POWer:REFerence <Reference>
Queries the medial threshold level of the overall pulse level. This level is used to define
the pulse width and pulse period.
Note: This parameter is only avalaible in time measurement mode and R&S NRP-Z81
power sensors.
Parameters:
<Reference> float
Range: 0.0 to 100.0
Increment: 0.01
*RST: 50.0
Example:
TRAC3:SWE:PULS:THR:POW:REF?
Queries the medial threshold level of trace 3.
Manual operation: See "Mesial" on page 267
:TRACe<ch>[:POWer]:SWEep:STATe <State>
Activates the selected trace.