List and sweep mode
R&S
®
SMA100B
188User Manual 1178.3834.02 ─ 09
Table 7-7: General settings dialog: Examples of RF Frequency and RF Combined Sweep dialogs
> "RF Frequency Sweep" > "RF Combined Sweep"
Settings
Sweep Type................................................................................................................ 188
State (RF frequency sweep)....................................................................................... 188
State (RF level sweep)................................................................................................189
State (RF frequency / level sweep).............................................................................189
State (LF frequency sweep)........................................................................................189
Current Frequency...................................................................................................... 189
Current Level...............................................................................................................189
Mode........................................................................................................................... 189
Retrace........................................................................................................................190
Shape..........................................................................................................................191
Spacing....................................................................................................................... 192
Sweep Time................................................................................................................ 192
Dwell Time...................................................................................................................192
Trigger Slope...............................................................................................................193
Show Connector..........................................................................................................193
Reset Sweep...............................................................................................................193
Execute Single Sweep................................................................................................ 193
Sweep Type
Selects the kind of the frequency sweep.
"Step"
Increases or decreases the frequency at each step.
To determine the frequency intervals, see Spacing.
"Ramp"
Provides a synthesized continuous analog frequency sweep with the
set Sweep Time.
Remote command:
[:SOURce<hw>]:SWEep:GENeration on page 701
State (RF frequency sweep)
Activates RF frequency sweep signal generation.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.