List and sweep mode
R&S
®
SMA100B
198User Manual 1178.3834.02 ─ 09
"High Dyn. Uninterrupted"
R&S SMAB-K724
Provides a linear output power that is uninterrupted over a wide
dynamic range.
Note: The R&S SMA100B supports this characteristic at frequencies
above 52 MHz. If you select the setting at lower frequencies
(52 MHz), the instrument reports a settings conflict.
"User"
Indicates that a setting has been modified in the expert mode. The
expert mode is a protected function, that requires protection level 2
password.
Remote command:
[:SOURce<hw>]:POWer:LBEHaviour on page 686
Sweep Level Range
Displays the RF level sweep range.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Remote command:
[:SOURce<hw>]:POWer:RANGe:MIN? on page 691
[:SOURce<hw>]:POWer:RANGe:MAX? on page 691
7.5.4 Output settings
Option: R&S SMAB-B28
Access:
1. Select "Sweep" > "RF Frequency Sweep".
2. Select "Output".
Settings
V/GHz / X-Axis............................................................................................................ 198
Marker / User.............................................................................................................. 199
V/GHz / X-Axis
Option: R&S SMAB-B28
Selects the mode the voltage is supplied depending on the frequency.
Sweep mode settings