The connector types listed in this table represent the common connectors provided by
Rohde & Schwarz. It is considered as general information and therefore can contain
connector types that do not apply to your instrument.
See Chapter 3.2.1.11, "RF 50 ", on page 41.
Table 3-1: Connector types and torque limits
Type Torque limit Nut opening
lb-Inch Nm Inch mm
N 13.3 1.5 3/4 20
SMA 5 0.56 5/16 8
3.5 mm 8 0.9 5/16 8
2.92 mm 8 0.9 5/16 8
2.4 mm 8 0.9 5/16 8
1.85 mm 8 0.9 5/16 8
1.0 mm 3 0.34 0.236 6
To prevent RF output switch-off
NOTICE! If you set a too high output level without a load connected to the instru-
ment, the reverse power can exceed a limit forcing the R&S SMA100B to switch off
the RF output.
Connect a load with sufficient return loss as given in the data sheet.
3.1.10 Connecting to Ref In/Ref Out
The connector is located on the rear panel.
To connect to Ref In/Ref Out (reference < 1 GHz)
For connection, the R&S SMA100B provides BNC connectors.
To connect the cable with the "Ref In"/"Ref Out" connector, proceed as follows:
a) Carefully align the connector of the cable and the "Ref In"/"Ref Out" connector
along a common axis.
b) Mate the connectors along the common axis until the male pin of the connector
of the cable engages with the female socket of the "Ref In"/"Ref Out" connec-
tor.
To connect to Ref In/Ref Out (reference = 1 GHz)
For connection, the R&S SMA100B provides SMA connectors.
1. Use a high-quality cable that matches the connector type.
See "Cable selection and electromagnetic interference (EMI)" on page 29.