Network operation and remote control
R&S
®
SMA100B
412User Manual 1178.3834.02 ─ 09
Changing the LAN configuration
This function is password-protected. Unlock the protection level 1 to access it.
Note: We recommend that you change the default password before connecting the
instrument to a network.
How to:
"Changing the default user password of the instrument" on page 373.
"Changing the default security password" on page 374.
See Chapter 12.4.4, "Password management", on page 370.
Ping client
The "Ping Client" page provides the ping utility to verify the connection between the
LXI-compliant instrument and another device.
The ping is initiated from the instrument. Using the ICMP echo request and echo reply
packets, this function checks whether the communication with a device via LAN is
working. Ping is useful for the diagnosis of IP network or router failures.
To initiate a ping at the instrument:
1. On the "Ping Client" page, enter the IP address of the host in the "Destination
Address" field (for example 10.113.1.151).
2. Select "Submit".
13.6.2.2 SCPI remote trace
The remote trace functionality allows you to trace input and output strings at the
remote control interface of the R&S SMA100B, see Chapter 13.8.8, "How to trace mes-
sages with the LXI web browser interface", on page 428.