Reference oscillator
R&S
®
SMA100B
296User Manual 1178.3834.02 ─ 09
Deactivate RF Output (if external reference is missing)..............................................296
External Reference Frequency................................................................................... 296
Variable Reference Frequency....................................................................................297
Synchronization Bandwidth.........................................................................................297
Nominal Synchronization Bandwidth...........................................................................297
Minimum Locking Range.............................................................................................297
External Tuning Active................................................................................................ 298
External Tuning Slope.................................................................................................298
Set to Default
Calls the default settings.
Remote command:
[:SOURce]:ROSCillator:PRESet on page 693
Source
Selects the reference frequency source.
"Internal"
Uses the internal reference oscillator, either with calibrated or a user-
defined Adjustment Value.
"External"
Uses an external reference signal.
Note: If the external reference is missing, the R&S SMA100B issues
a warning message and indicates the icon (external reference
missing).
To set the frequency of the external reference, see "External Refer-
ence Frequency" on page 296.
Remote command:
[:SOURce]:ROSCillator:SOURce on page 693
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Deactivate RF Output (if external reference is missing)
Turns off the RF output when the external reference signal is selected, but no signal is
supplied.
This function prevents that no improper RF signal due to the missing external refer-
ence signal is used for measurements. A message indicates that the RF output is
deactivated.
Remote command:
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe] on page 694
External Reference Frequency
Selects the frequency of the external reference signal.
Set the reference output frequency with the parameters in the Reference output set-
tings dialog.
"10
MHz"
Selects 10 MHz for the external reference frequency signal.
Reference frequency settings