288User Manual 1178.3834.02 ─ 09
11. Press "Start Cont" to start the measurement.
8.4.4.9 How to set up a power sweep measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 8.4, "Using power sensors", on page 233.
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Power".
4. Configure the measurement:
a) Set "Min = -20 dBm".
b) Set "Max = 30 dBm".
c) Set "Steps = 200".
5. Select the "Configure Diagram" tab.
Ho to: see Chapter 8.4.4.11, "How to configure the power analysis diagram",
on page 289.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. If you want to configure the frequency settings of the sensor differently than the
generator settings, enable "Use Separate Frequ.".
a) Set "1 MHz".
9. In the "Next Window List" tab, enable the "Standard View".
10. Close the "Configure" dialog.
11. Press "Start Cont" to start the measurement.
8.4.4.10 How to set up a pulse measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 8.4, "Using power sensors", on page 233.
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Time".
4. Configure the measurement:
a) Set "Min = -5 us".
b) Set "Max = 45 us".
c) Set "Steps = 200".