Improving level performance
R&S
®
SMA100B
287User Manual 1178.3834.02 ─ 09
a) To change the default directory the file is saved in, select " Path" and define a
path and a filename. For example, select the default directory /var/user.
b) If necessary, disable or change some of the parameters in the "Automatic Nam-
ing Settings".
c) Close the "Save NRP-Z Analysis Options" dialog.
8. In the "Save NRP-Z Analysis" dialog, select "Save".
The instrument saves a hardcopy of the current instrument display as a *.jpg file.
The filename is automatically created.
9. To print the hardcopy, connect the instrument to a LAN and:
a) Transfer the file to a remote computer as described in Chapter 11.9, "How to
transfer files from and to the instrument", on page 330.
b) On the remote computer, navigate through the file system.
c) Print the selected file.
For more information, refer to the online help of the operating system.
8.4.4.8 How to set up a frequency sweep measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 8.4, "Using power sensors", on page 233.
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Frequency".
4. Configure the measurement:
a) Set "Min = 8 KHz".
b) Set "Max = 6 GHz".
c) Set "Steps = 200".
5. Select the "Configure Diagram" tab.
Ho to: see Chapter 8.4.4.11, "How to configure the power analysis diagram",
on page 289.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. If you want to configure the frequency settings of the sensor differently than the
generator settings, enable "Use Separate Frequ.".
a) Set "Min = 1 MHz".
b) Set "Max =10 MHz".
9. In the "Next Window List" tab, enable the "Standard View".
10. Close the "Configure" dialog.
Using power sensors