EasyManua.ls Logo

R&S SMA100B - How to Calibrate the Power Level with an R&S NRP Power Sensor

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Improving level performance
R&S
®
SMA100B
290User Manual 1178.3834.02 ─ 09
8.5 How to calibrate the power level with an R&S NRP
power sensor
Using a R&S NRP power sensor and the user correction function of the instrument,
you can compensate external frequency responses. The R&S SMA100B utilizes the
readings of the power sensor and creates a correction value table for controlling the
output level during operation.
To connect the R&S NRP to the R&S SMA100B
1. Connect the power sensor.
See Chapter 8.4.1, "Connecting R&S NRP power sensors to the R&S SMA100B",
on page 233.
2. Select "Clk Syn/Power Sens" > "NRP Sensor Mapping".
3. If the sensor is not detected automatically, select "Scan > Start".
The instrument scans the network and the USB connections for connected sensors
and lists all detected R&S NRP sensors in the mapping table.
4. In the "Mapping" column, assign the sensor to a sensor channel, see Chap-
ter 8.4.2, "NRP sensor mapping", on page 234.
5. Close the dialog.
To configure and calibrate the R&S NRP in the R&S SMA100B
Provided the power sensor is connected to the R&S SMA100B and is assigned to a
sensor channel, we recommend that you calibrate and configure the power sensor in
the "NRP Power Viewer" dialog.
1. Select "Level" > "RF ON > Off"
2. Select "Clk Syn/Power Sens" > "NRP Power Viewer".
3. If the R&S SMA100B has detected several connected power sensors, select the
tab of the sensor you want to use.
4. Select "Zero" to start zeroing of the sensor.
Note: Always turn the RF power off or disconnect the sensor from RF before zero-
ing, since the function calibrates the power sensor at zero signal power.
The zeroing process takes a few seconds, depending on the power sensor type.
5. Configure additional parameters for the selected sensor as required.
How to calibrate the power level with an R&S NRP power sensor

Table of Contents

Other manuals for R&S SMA100B

Related product manuals