Maintenance, storage and disposal
R&S
®
SMA100B
762User Manual 1178.3834.02 ─ 09
Minimum power rating required for internal adjustments
The following tables list the minimum power rating of external 50 Ω termination that is
required to execute internal adjustments. The minimum power rating depends on the
output power options, the instrument is equipped with.
Table 17-1: Standard (no option) or high output power (R&S
SMAB-B35/-B37/-B39/-K31/-K33/)
Frequency option R&S SMAB-B103/../-B120 R&S SMAB-B131/../-B167 R&S SMAB-B131/../-B167
Firmware version not relevant <4.60.112.29SP5 4.60.112.29SP5
Min. power rating 1 W (30 dBm) 1 W (30 dBm) no termination required
Table 17-2: Ultra high output power (R&S
SMAB-B32/-B34/-K36/-K38/-K40)
Frequency option R&S SMAB-B103/../-B120 R&S SMAB-B131/../-B167 R&S SMAB-B131/../-B167
Firmware version not relevant <4.60.112.29SP5 4.60.112.29SP5
Min. power rating 2 W (33 dBm) 2 W (33 dBm) no termination required
Table 17-3: Super ultra high output power (R&S
SMAB-B36S)
Frequency option not relevant
Firmware version not relevant
Min. power rating 2 W (33 dBm)
17.3.1.1 Internal adjustments settings
Access:
Select "System Config > Setup > General > Internal Adjustments".
In this dialog, you can perform internal calibration routines and get information on
the last performed calibration.
Performing maintenance tasks