Maintenance, storage and disposal
R&S
®
SMA100B
763User Manual 1178.3834.02 ─ 09
The "Temperature Offset" indicates the deviation of the current temperature of the
instrument, compared to the temperature of the last adjustment.
The remote commands required to define these settings are described in Chapter 14.6,
"CALibration subsystem", on page 467.
How to: See Chapter 17.3.1.2, "How to use the internal adjustments", on page 764.
Adjust All
Performs all available internal calibration routines of the instrument.
How to: See Chapter 17.3.1.2, "How to use the internal adjustments", on page 764.
Remote command:
:CALibration:ALL[:MEASure]? on page 467
Last Full Adjustment
Displays the date of the last fully performed adjustment.
Remote command:
:CALibration<hw>:ALL:DATE? on page 467
Time
Displays the elapsed time since the last full adjustment.
Remote command:
:CALibration<hw>:ALL:TIME? on page 468
Temperature Offset
Displays the temperature difference, comparing the temperature of the last adjustment
to the current instrument temperature.
A green checkmark indícates that the offset is within the permitted range. If the temper-
ature deviates more than ±5 K, the instrument indicates a warning icon.
Remote command:
:CALibration<hw>:ALL:TEMP? on page 468
Information
Displays information to the current adjustment state.
Remote command:
:CALibration<hw>:ALL:INFormation? on page 468
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:CONTinueonerror on page 469
Performing maintenance tasks