Remote control commands
R&S
®
SMA100B
468User Manual 1178.3834.02 ─ 09
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Date> string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Full Adjustment" on page 763
:CALibration<hw>:ALL:INFormation?
Queries the current state of the internal adjustment.
Return values:
<CalInfoText> string
Example:
CAL:ALL:INF?
"Instrument is calibrated, no adjustment required."
"UNCAL, instrument is warming up."
"UNCAL, Please perform full adjustment after warming up."
"UNCAL, Please perform full adjustment."
Usage: Query only
Manual operation: See "Information" on page 763
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Temperature> string
Example:
CALibration:ALL:TEMP?
// "+12.00 K"
Usage: Query only
Manual operation: See "Temperature Offset" on page 763
:CALibration<hw>:ALL:TIME?
Queries the time elapsed since the last full adjustment.
Return values:
<Time> string
CALibration subsystem