Loading...
Improving level performance
R&S
®
SMA100B
272User Manual 1178.3834.02 ─ 09
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:REFerence? on page 540
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:REFerence:DISPlay:
ANNotation[:STATe] on page 541
Trigger settings
Access:
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Time".
4. Select "Configure Time Mode"
5. Select the "Trigger" tab.
Sensor - Power Analysis
Selects the power sensor to be set if more than one sensor is connected to the instru-
ment.
Remote command:
n.a.
In remote control, the sensor is selected via the numeric suffix in the sense key word of
the command, for example SENSe2:POWer:SWEep:….
Mode - Power Analysis
Selects if the measurement is free running, or starts only after an internal or external
trigger event.
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:SOURce on page 511