Remote control commands
R&S
®
SMA100B
511User Manual 1178.3834.02 ─ 09
Parameters:
<Hysteresis> float
Range: 0 to 10
Increment: 0.001
*RST: 0.5
Example:
SENS1:SWE:TIME:SENS:TRIG:HYST 0.5 dB
The hysteresis is 0.5 dB
Manual operation: See "Hysteresis - Power Analysis" on page 273
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:LEVel
<Level>
Sets the trigger threshold.
Parameters:
<Level> float
Range: -200 to 100
Increment: 0.001
*RST: 1
Example:
SENS1:SWE:TIME:SENS:TRIG:LEV -20 dBm
Sets the trigger level to -20 dBm.
Manual operation: See "Level - Power Analysis " on page 273
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:SLOPe <TriggerSlope>
Sets the polarity of the active slope for the trigger signals.
Parameters:
<TriggerSlope> POSitive | NEGative
*RST: POSitive
Example:
SENS1:SWE:TIME:TRIG:SLOP POS
The positive edge of a trigger signal is active.
Options: Option R&S SMAB-K28
Manual operation: See "Slope - Power Analysis" on page 273
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:TRIGger:SOURce <Source>
Selects if the measurement is free running (FREE) or starts only after a trigger event.
The trigger can be applied internally or externally.
Parameters:
<Source> FREE | AUTO | INTernal | EXTernal
*RST: AUTO
Example:
SENS1:SWE:TIME:SENS:TRIG:SOUR FREE
The power versus time measurement is performed free running
Power sensor measurement subsystems