Reference oscillator
R&S
®
SMA100B
299User Manual 1178.3834.02 ─ 09
"Off"
Deactivates the reference signal output.
"10 MHz"
Derives a signal with 10 MHz frequency from the internal reference
oscillator and provides this signal at the output.
"100 MHz"
Option: R&S SMAB-K703
Provides the 100 MHz reference frequency signal at the output.
"1 GHz"
Option: R&S SMAB-K703
Provides the 1 GHz reference frequency signal at the output.
"Input Signal (loop through)"
Option: R&S SMAB-K703/-K704
Passes the external reference frequency to the output directly.
Remote command:
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE on page 695
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE on page 696
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
9.4 Adjustment settings
Access:
1. Select "Frequency" > "Reference Frequency".
2. Select "Adjustment".
Settings:
Adjustment Active....................................................................................................... 299
Adjustment Value........................................................................................................ 300
Adjustment Active
Selects the adjustment mode.
"Off"
Uses the calibrated internal reference frequency.
"On"
Allows you to apply a deviation to the internal reference frequency,
according to your requirements.
Enter the value in the Adjustment Value field.
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] on page 697
Adjustment settings