Remote control commands
R&S
®
SMA100B
695User Manual 1178.3834.02 ─ 09
For more information, see data sheet.
Parameters:
<SBandwidth> WIDE | NARRow
NARRow
The synchronization bandwidth is a few Hz.
WIDE
Uses the widest possible synchronization bandwidth.
*RST: n.a. (factory preset)
Example: See Example "Configuring the reference oscillator" on page 692.
Manual operation: See "Synchronization Bandwidth" on page 297
[:SOURce]:ROSCillator:EXTernal:MLRange?
Queries the minimum locking range for the selected external reference frequency.
Depending on the RF hardware version, and the installed options, the minimum locking
range vaies.
For more information, see data sheet.
Return values:
<MinLockRange> string
Example:
SOUR:ROSC:EXT:MLR?
Usage: Query only
Manual operation: See "Minimum Locking Range" on page 297
[:SOURce]:ROSCillator:EXTernal:NSBandwidth?
Queries the nominal synchronization bandwidth for the selected external reference fre-
quency and synchronization bandwidth.
Return values:
<NomBandwidth> string
Example:
SOUR:ROSC:EXT:NSB?
Usage: Query only
Manual operation: See "Nominal Synchronization Bandwidth" on page 297
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE <OutpFreqMode>
Sets the output reference frequency.
Parameters:
<OutpFreqMode> DER10M | DER100M | OFF | LOOPthrough
OFF
Disables the output.
SOURce subsystem