Example: See [:SOURce<hw>]:INPut:MODext:COUPling<ch>
on page 651.
Manual operation: See "Impedance" on page 118
[:SOURce]:INPut:TRIGger:SLOPe <Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example:
INP:TRIG:SLOP NEG
Activates the falling slope of the external trigger signal at the
trigger input.
Manual operation: See " Trigger Slope" on page 193
14.16.6 SOURce:LFOutput subsystem
The SOURce:LFOutput subsystem contains the commands for setting the LF signal
source in CW and Sweep mode and for analog modulation.
Example: Setup an LF sweep
The following example shows a command sequence to set up an LF sweep.
// Reset the instrument to start from an initial state
*RST; *CLS
// Set the trigger mode, the sweep mode and the sweep range
TRIGger1:LFFSweep:SOURce SINGle
SOURce1:LFOutput1:SWEep:FREQuency:MODE AUTO
SOURce1:LFOutput1:FREQuency:STARt 1 kHz
SOURce1:LFOutput1:FREQuency:STOP 7 kHz
// Select linear spacing
// Select the waveform shape for the frequency sweep cycle
// Set the step width and the dwell time.
SOURce1:LFOutput1:SWEep:FREQuency:SPACing LINear
SOURce1:LFOutput1:SWEep:FREQuency:SHAPe SAWtooth
SOURce1:LFOutput1:SWEep:FREQuency:STEP:LINear 100 Hz
SOURce1:LFOutput1:SWEep:FREQuency:DWELl 150 ms
// Alternatively to the step width set the number of steps
SOURce1:LFOutput1:SWEep:FREQuency:POINts 61
// Activate change to start frequency while waiting for next trigger
// Prerequisites: sweep mode single and sweep waveform sawtooth
SOURce1:LFOutput:SWEep:FREQuency:RETRace 1