Improving level performance
R&S
®
SMA100B
289User Manual 1178.3834.02 ─ 09
5. Select the "Configure Diagram" tab.
How to: see Chapter 8.4.4.11, "How to configure the power analysis diagram",
on page 289.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. In the "Next Window List" tab, enable the "Pulse Data View".
9. Close the "Configure" dialog.
10. In the diagram window, open the context-sensitive menu.
11. Select "Trace/Pulse Data Notifications/ Trigger".
12. Select the "Trace" tab.
13. Set "State > On".
14. Set the thresholds:
a) Set "Base > Voltage Related".
b) Set "Distal = 90%".
c) "Mesial = 50%".
d) "Proximal = 10%".
15. Select the "Pulse Data Notifications" tab.
16. Enable up to six parameters, for example:
● "Duty Cycle"
● "Pulse Width"
● "Pulse Period"
● "Off Time"
● "Start Time"
17. Close the "Trace/Pulse Data Notifications/ Trigger" dialog.
18. Press "Start Cont" to start the measurement.
8.4.4.11 How to configure the power analysis diagram
1. Select "Clk Syn/Power Sens" > "NRP-Z Power Analysis".
2. Select "Configure" > "Configure Diagram".
3. Set "Max Level = 30 dBm".
4. Set "Min Level = -40 dBm".
5. Select "Auto Scale > Expand Only (Coarse)".
6. Enable "Show Grid".
7. Set "Background Color > Black."
Using power sensors